User Tools

Site Tools


characterization_machine_documentation

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
characterization_machine_documentation [2015/06/16 17:17]
sgneitin
characterization_machine_documentation [2015/07/21 12:46] (current)
sgneitin ↷ Links adapted because of a move operation
Line 29: Line 29:
 ^                    ^  Red      ^  Green    ^  Blue     ^ ^                    ^  Red      ^  Green    ^  Blue     ^
 |  Laser Voltage ​    ​| ​ 5V       ​| ​ 5.5V     ​| ​ 6.5V     | |  Laser Voltage ​    ​| ​ 5V       ​| ​ 5.5V     ​| ​ 6.5V     |
-|  Angular Reading ​  ​| ​ 354°     ​|  ​356°     |  ​358° ​    | +|  Angular Reading ​  ​| ​ 354°     ​|  ​357°     |    ​5° ​    | 
-|  Center Frequency ​ |  ​425 MHz  |  ​490 MHz  |  ​570 MHz  | +|  Center Frequency ​ |  ​450 MHz  |  ​480 MHz  |  ​550 MHz  |
  
 ==Alignment== ==Alignment==
 Now that you have mounted the device you are ready to align the machine. Using the table in the section before do the following: Now that you have mounted the device you are ready to align the machine. Using the table in the section before do the following:
  
-  - Connect the RF cable to the Transducer you wish to test.+  - Connect the RF cable to the Transducer you wish to test.  ​{{ ​ ::​a7_070.jpg?​200| ​ }}
   - Power on the Laser to the appropriate voltage.   - Power on the Laser to the appropriate voltage.
   - Check the polarization (adjust for least light transmitted through a horizontal polarizor).   - Check the polarization (adjust for least light transmitted through a horizontal polarizor).
   - With the adjustable aperture all the way open rotate the sample and position it to the correct angular reading.   - With the adjustable aperture all the way open rotate the sample and position it to the correct angular reading.
-  - Adjust the XY stage and rotation stage until waveguiding in the desired mode is achieved. It should look like this:  {{  ::a7_070.jpg?​100| ​ }}  +  - Adjust the XY stage and rotation stage until waveguiding in the desired mode is achieved. It should look like the picture on the right.  ​
   - Set RF generator to the correct center frequency. (Verify that is is outputting a signal and that the amplifier is on)   - Set RF generator to the correct center frequency. (Verify that is is outputting a signal and that the amplifier is on)
   - Move the platform vertically until you see deflected light.   - Move the platform vertically until you see deflected light.
Line 60: Line 59:
  
 Good results look like this:  Good results look like this: 
-{{:​aom-a1-green-6.10.2015-8.55am-graph.jpeg?​300|}}+{{:samples:a:​aom-a1-green-6.10.2015-8.55am-graph.jpeg?​300|}}
  
 Once you have finished testing all colors and modes run it through the Matlab code to compare results. Once you have finished testing all colors and modes run it through the Matlab code to compare results.
characterization_machine_documentation.1434496637.txt.gz · Last modified: 2015/06/16 17:17 by sgneitin